PRICES include / exclude VAT
Homepage>BS Standards>77 METALLURGY>77.040 Testing of metals>77.040.30 Chemical analysis of metals>BS EN 10315:2006 Routine method for analysis of high alloy steel by X-ray fluorescence spectrometry (XRF) by using a near by technique
Sponsored link
immediate downloadReleased: 2009-09-30
BS EN 10315:2006 Routine method for analysis of high alloy steel by X-ray fluorescence spectrometry (XRF) by using a near by technique

BS EN 10315:2006

Routine method for analysis of high alloy steel by X-ray fluorescence spectrometry (XRF) by using a near by technique

Format
Availability
Price and currency
English Secure PDF
Immediate download
249.60 EUR
You can read the standard for 1 hour. More information in the category: E-reading
Reading the standard
for 1 hour
24.96 EUR
You can read the standard for 24 hours. More information in the category: E-reading
Reading the standard
for 24 hours
74.88 EUR
English Hardcopy
In stock
249.60 EUR
Standard number:BS EN 10315:2006
Pages:32
Released:2009-09-30
ISBN:978 0 580 68841 6
Status:Standard
DESCRIPTION

BS EN 10315:2006


This standard BS EN 10315:2006 Routine method for analysis of high alloy steel by X-ray fluorescence spectrometry (XRF) by using a near by technique is classified in these ICS categories:
  • 77.140.20 Stainless steels
  • 77.040.30 Chemical analysis of metals

This European Standard specifies a procedure on how to improve the performance of a routine XRF method, already in use for analysis of high alloy steels, by using a ”near by technique”.

The ”near by technique” requires at least one target sample (preferable a CRM) of a similar composition as the unknown sample.

The method is applicable to elements within the concentration ranges according to Table 1:

Table 1 — Concentration ranges

Element Concentration range, % (m/m) a
Si 0,05 to 1,5
Mn 0,05 to 5,0
P 0,005 to 0,035
Cr 10 to 25
Ni 0,1 to 30
Mo 0,1 to 6,5
Cu 0,02 to 1,5
Co 0,015 to 0,30
V 0,015 to 0,15
Ti 0,015 to 0,50
Nb 0,05 to 1,0

a The concentration ranges specified, represents those ranges studied during the precision test. The procedure has the potential to be used outside those ranges but it needs to be validated by each laboratory in every case.

The method is applicable to analysis of either chill-cast or wrought samples having a diameter of at least 25 mm and with a carbon concentration of less than 0,3 % (see NOTE). Other elements should have a concentration below 0,2 %.

NOTE High carbon concentrations, in combination with high Mo and Cr concentrations, could have undesirable structural effects on the sample and could affect the determination of phosphorus and chromium, in particular.

Matrix effects exist between the elements listed. To compensate for those inter-element effects, mathematical corrections shall be applied. A variety of computer programs for corrections is commonly used and included in the software package from the manufacturers.